SpeechPathology.com Phone: 800-242-5183


Fall in love with your career at Stepping Stones Group - Apply Now

Fitting Techniques for the Pediatric Cochlear Implant Patient

Fitting Techniques for the Pediatric Cochlear Implant Patient
Aimee Gross
May 12, 2003
Share:

 

Introduction
A cochlear implant is a prosthetic device designed to provide hearing to patients with severe to profound sensorineural hearing loss. A portion of the device, referred to as the ''implant package,'' is surgically placed in the mastoid portion of the temporal bone. An electrode array extending from the implant package is inserted into the cochlea. The remaining portions of the device are worn externally and include a microphone, speech processor, cable, and transmitting coil. Figure 1 illustrates the MED-EL Cochlear Implant System and TEMPO+ Speech Processor.



Sound is collected by the microphone (1) and sent to electronic components within the speech processor (2). The speech processor analyzes the input signal (sound) and converts it into an electronic code. This code travels along a cable (3) to the transmitting coil (4) and is sent across the skin via frequency modulated (FM) radio waves to the implant package (5). Based on characteristics of the code transmitted to the internal device, electrode contacts within the cochlea (6) provide electrical stimulation to the spiral ganglion cells and dendrites extending into the modiolus. Electrical impulses then travel along the auditory nerve (7) and ascending auditory pathways to the brain.

The characteristics and code transmitted by a cochlear implant are dependent, not only on the components of the device itself, but also on the way in which the device is fit. Fitting of a cochlear implant, also referred to as ''programming'' or ''mapping,'' creates a set of instructions (code) that defines the specific characteristics used to stimulate the electrodes of the implanted array. This set of instructions is referred to as the patient's ''program'' or ''map.'' Implant programming is performed by an audiologist with specialized training in the field of cochlear implants. The audiologist uses interactive software and computer hardware to create individualized patient programs that are stored on the hard drive of the computer and downloaded to the memory of the patient's speech processor.

With advances in cochlear implant technology, the fitting process has become more complex. Today's implants offer a number of sophisticated parameters that can be manipulated to improve sound quality and speech understanding. All cochlear implants require at least two levels to be set for each stimulating electrode. These measurements are referred to as the Threshold level (abbreviated as THR or T) and the Most Comfortable Loudness level (abbreviated as MCL, M, or C). Exact terminology and abbreviations are device specific, but the general purpose of these measures is common across all cochlear implants.

 

Aimee Gross



Related Courses

Treatment Approach Considerations for School-Aged Children with Speech Sound Disorders
Presented by Kathryn Cabbage, PhD, CCC-SLP
Video
Course: #9472Level: Intermediate1 Hour
This course will address the theoretical underpinnings and research base related to differential diagnosis and treatment of articulation and phonological deficits in children with speech sound disorders. Special considerations for how to tailor evaluation and intervention to meet the needs of school-age children will be discussed.

The Ripple Effect of Stuttering: A Community-Based Approach
Presented by Craig Coleman, MA, CCC-SLP, BCS-F, ASHA Fellow, Mary Weidner, PhD, CCC-SLP
Video
Course: #9217Level: Intermediate2 Hours
This is Part 2 of a four-part series. The stuttering experience has a ripple effect that extends far beyond the child who stutters. Parents, teachers, peers, and others must possess both knowledge and skills to best support children who stutter. This course will highlight new clinical tools and resources to provide a community-based treatment approach for stuttering. (Part 1 - Course 9278, Part 3 - Course 9301, Part 4 - Course 9304)

Behavioral Frameworks for Dementia Management
Presented by Mary Beth Mason, PhD, CCC-SLP, Robert W. Serianni, MS, CCC-SLP, FNAP
Video
Course: #9473Level: Intermediate1 Hour
This course will focus on cognitive-communication intervention strategies for various dementia presentations and will provide a review of evidence-based treatment. Behavioral frameworks along with their rationales will be introduced and applied across several dementia types and mild, moderate and severe levels of impairment.

20Q: Infection Control Strategies for SLPs
Presented by A.U. Bankaitis Smith, PhD
Text
Course: #9729Level: Intermediate1 Hour
Speech-language pathologists are expected by policy authorities to apply appropriate measures to protect patients, co-workers and themselves in clinical situations that may expose individuals to infectious microbes. This article provides practical guidelines for implementing infection control principles within the context of the COVID-19 pandemic, including discussion of personal protective equipment (PPE) and disinfecting and cleaning products.

Thickened Liquids in Clinical Practice: The Plot “Thickens”
Presented by Angela Mansolillo, MA, CCC-SLP, BCS-S
Video
Course: #10497Level: Intermediate1 Hour
Clinicians who utilize thickened liquids in their clinical practice are aware of their benefits, but what about the risks and contraindications? Advantages and disadvantages of thickened liquids are reviewed in this course with a focus on clinical outcomes, including impacts on medication administration, lung health, and hydration. Product types are evaluated to facilitate appropriate choices for individual clients.

Our site uses cookies to improve your experience. By using our site, you agree to our Privacy Policy.